Specifications
GM Input: BNC connector. Accepts standard
Geiger Counters. Common HV.
SCINT Input: MHV connector. Anode signal
from scintillation detector. Common HV.
High voltage: 0 to +1200 volts,
stabilized. Digitally selectable in 10 volt increments.
Display: Computer display. See STX
software for more information here.
Modes: Count for preset time, counts/min.,
counts/sec., set alarm level 0-999999, set high voltage 0-1200v. Auto sensing
serial or USB connection.
Audio: Beeps on input counts. Continuous
tone if count rate exceeds preset level.
Data Links: USB connector for PC and
Macintosh. DB-9 connectors accept RS-232 serial cable for computers not equipped
with USB.
Power: Input 9 volt DC, at 500mA from AC
line/charger. Specify 110-120, 220-240 VAC at time of order.
Dimensions: 12 in. W, x 8 in. H, x 4.5 in.
D.