Pen Dosimeter Specifications   PenDosimeters

Energy Dependence
±10% maximum change in sensitivity for x and gamma ray energies from 16 keV to 8 MeV.
Electrical Leakage
In the absence of radiation, the dosimeters leakage or self discharge is less than 0.5% of full scale in 24 hours at 50° C. At 20° C dosimeters will read background radiation. After gross over exposure totaling up to 2000 R the leakage is less than 5% of full scale in 48 hours at 20° C. Thereafter, the leakage will decrease to original values.
Charging Voltage
Dosimeters are capable of being charged to "zero" with 140 to 195 volts.
Operational Characteristics
A maximum change in sensitivity of ±10% may occur under any probable combination of the following:
A) Temperature range below -20°C (-4°F) or above +50°C (122°F)
B) Greater than 50,000 foot altitude for one month
C) Relative humidity greater than 95% Highly resistant to shock and vibration.
Accuracy
±10% of true dose for Cs-137 or Co-60 gamma.
Rate Response
Dosimeters are dose rate independent.

Enclosure Material
Barrel - carbon fiber-filled, high-strength, electrically-conducting plastic with metal clip.
Weight
Approx. 19 grams (0.7 oz.)
Size
PEN- 200, 500
2mSv, 5mSv Length: 11 cm (4.48 inches)
Diameter: 1.5 cm (0.6 inches)
PEN- 2, 5, 20 Length: 11.5 cm (4.52 inches)
Diameter: 1.5 cm (0.6 inches)
Options
Protective sapphire lens

 